When the Zoom dialog box appears press the reset zoom button. Touch or click the Zoom (Z1) trace annotation box. The grayed out areas in the Address Setup and Data Pattern setup will activate.Įnter the value 4C into the Address field in the Address Setup.Įnter the value 65 into the Data Value field in the Data Pattern Setup. Press the button that is labeled Addr+Data in the Trigger type section of the I2C trigger setup. We will setup a trigger based on the I2C address and data content. There are 8 trigger types to choose from.

#Wavesurfer mxs serial

Having checked the Link to Trigger box on the serial decode setup you will find that the basic configuration of the signals sources has been automatically transferred to the trigger setup. This will cause the trigger dialog box to appear.

  • Set Time/Division: 20 ms, Select Set Maximum Memory and set Maximum Sample Points to 13MS.
  • wavesurfer mxs

    This will bring up the Timebase dialog box.

  • Touch or click the Timebase annotation box on the display of the oscilloscope.
  • Connect the ground lead of the probe to the ground connection on the WS-100 as shown in Figure 1.
  • Connect one passive probe to Ch2 of the scope and Ch2 of the WS 100.
  • Connect one passive probe to Ch1 of the scope and Ch1 of the WS 100.
  • If the Found New Hardware Wizard appears press the cancel button.
  • Connect the WS 100 to a front panel USB connector.
  • Press the Normal button in the Trigger area on the oscilloscope’s front panel.
  • Set up the trigger as follows: Edge-Trigger, Source :C1, Coupling : DC, Slope: Positive, Trigger Level: 2.5 V
  • Touch or click the Trigger annotation box on the display of the oscilloscope.
  • wavesurfer mxs

    Using the C1 dialog box of the front panel vertical scale control, set the channel 1 vertical scale to 1 V/division and the vertical offset to -3V.Verify that the input coupling on both channels 1 and 2 is DC 1 M: Touch or click the channel 1 annotation box>touch or click on the coupling field >select DC 1M.Recall the default setup: File pull down > Recall Setup> Recall Default.The Signal Outputs Available from the WS 100 Demonstration Signal Sourceĭisplays shown in the tutorial are based on the following initial setup on a WaveRunner 6 Zi scope: In this tutorial we will use selection S2 (I2C). Table 1 summarizes all the available outputs. There are two sets of duplicate output connections on the WS 100 which are marked Ch1, CH2, and Trig. The WS 100 signal source is a USB powered circuit board which has 6 sets of output waveforms.

    #Wavesurfer mxs how to

    Note: Although a specialized signal source is used in the following tutorial the explanations of how to set up measurements and interpret the results should be useful to WaveSurfer users who have other signal sources. WaveSource 100 demonstration signal source

    #Wavesurfer mxs series

    Equipment Requirements:Ī LeCroy WaveSurfer Xs-B series oscilloscope with the I2C option keycode installed. This tutorial will get you familiar with the LeCroy I2C Decode & Trigger Option. The slave responds with an acknowledge bit (ACK). Data transfer is initiated by the master with a start bit followed by the address of the desired slave and a read/write bit signaling that it wants to read from or write to the slave. Slaves are addressed using unique address codes which can be in a 7 or 10 bit format, multiple masters can be used in a single I 2C system, each with a unique address. I 2C uses two bidirectional lines with clock (SCL) and data (SDA) signals. I 2C is a master/slave synchronous communications system developed by Phillips with rates of up to 3.4 Mb/s. Capture digital signals up to 500 MHz with up to 10 Mpts/Ch memory, 2 GS/s, and 18 or 36 channels.Inter Integrated Circuit (Inter IC or I 2C) The 10 Mpts/Ch memory will allow you to maintain that sample rate for long captures of up to 2 ms or even longer for slower signals.Īdd high-performance mixed signal capability to a WaveSurfer MXs. With 5 GS/s on all four channels you will be able to capture your fastest signals.

    #Wavesurfer mxs plus

    On top of the great feature set and easy to use principles of the WaveSurfer Xs the WaveSurfer MXs provides outstanding specs, advanced triggers plus great math and measurement tools. The big display (but small footprint), simplified front panel, and graphical touch screen provide a friendly user interface for making measurements.

    wavesurfer mxs

    The WaveSurfer Xs is designed for fast validation and debug.